The optional white tray requires calibration before it can be used for the first time.
Note: If you subsequently obtain another white tray, Bio-Rad recommends that you calibrate the new tray.
When the imaging system detects that a white tray has been inserted for the first time, a message indicates that the tray is not calibrated and lists steps to calibrate it.
| 1. | Complete the steps listed in the message box. |
| 2. | Tap Calibrate. |
The system starts calibrating the white tray. A progress bar monitors the process.
A message notifies you when calibration is complete.
| 3. | Tap OK to close the message. |